Skip to main content

Table 2 Characteristics of the neonatal oxygenators [20]

From: Recent advancements in pediatric cardiopulmonary bypass technology for better outcomes of pediatric cardiac surgery

Name

Brand

Integrated ALF

Maximum blood flow (L/min

Priming volume (mL)

Surface area (m2)

Minimum-level in the reservoir (mL)

Biocoating

Quadrox-i Neonatal

Maquet Getinge

Yes

1.5

40

0.38

15

SOFTLINE (polymer coating) BIOLINE (heparin coating)

Capiox FX 0.5

Terumo

Yes

1.5

43

0.50

15

XcoatingTM (amphiphilic polymer)

KIDS D100

LivaNova

No

0.7

31

0.22

10

Phosphorylcholine coating

Affinity Pixie

Medtronic

No

2

48

0.48

20

CORTIVATM (heparin coating) BALANCE® (hydrophilic polymer)

HILITE 1000

Medos

No

1

57

0.39

50

Rheoparin® (heparin coating)

  1. Proper permission has been taken to reproduce this table from original author